Fisher And Paykel Healthcare Mr850 User Manual

  

Mr850 respiratory humidifier, Specifications, Warning

Advertising
Fisher paykel humidifier manual

SPECIFICATIONS

Dimensions:

The MR850 can be used for the full range of Fisher & Paykel Healthcare therapies across the F&P Respiratory Care Continuum™ One button selects optimal temperature and humidity levels for adult, pediatric and neonatal patients Easy set up with simple controls and an easy to read display Clinically based alarm system. See the MR850 technical manual for more detailed instructions on cleaning. Ensure that invasive mode is set for patients that have bypassed airways. The use of breathing circuits, chambers or other accessories which are not approved by Fisher & Paykel Healthcare may impair performance or compromise safety.

140 mm x 173 mm x 135 mm (without chamber fitted)

Weight:

2.8 kg (no chamber fitted), 3.1 kg (chamber fitted and filled with water)

Supply Frequency:

50/60 Hz Supply Voltage: 115 V~

Supply Current: 2.0 A max at 115 V~

Heater Plate:

150 W

Heater Wire:

22 V~, 2.73 A, 60 W, 50/60 Hz

Heater Plate Over-temperature Cutout: 118 ± 6 °C

Temperature Control Settings
Invasive Mode:

Chamber outlet: 35.5 - 37 °C, Airway 35 - 40 °C

Noninvasive Mode:

Chamber outlet: 31 °C, Airway 28 - 34 °C

Display:

Three digit 14 mm 7 segment LED

Fisher Paykel Humidifier Manual

Range:

10 - 70 °C, Accuracy: ± 0.3 °C (in 25 - 45 °C temperature range)

Alarm Parameters
High Humidity Alarm: An immediate, audible alarm at a displayed temperature of 41 °C or if the

airway temperature exceeds 43 °C

Low Humidity Alarm:

An audible alarm between 10 minutes @ 29.5 °C, and 60 minutes @ 34.5 °C
(Invasive Mode only)

Fisher And Paykel Healthcare Mr850 User Manual

Sound Pressure Level: Alarms exceed 50 dBA @ 1 m

Performance
Recommended ambient temperature range: 18 - 26 °C
CAUTION: If operating in ambient temperatures outside the recommended range, consult your local
Fisher & Paykel Healthcare representative.
Recommended Flow Range: Invasive Mode: up to 60 L/min, Noninvasive Mode: up to 120 L/min

Refer to breathing circuit specifications for minimum flow

Humidity Performance:

Invasive Mode: > 33 mg/L, Noninvasive Mode: >10 mg/L

Maximum Operating Pressure: Refer to chamber and breathing circuit specifications
Warm-up time:

Less than 30 minutes

Symbol Definitions

Type BF

ATTENTION

Consult

accompanying

documents

Alternating

Current

Drip

Proof

CAUTION

Hot surfaces

may exceed

85 °C

Respiratory Humidifier Classified by

Underwriters Laboratories Inc. with respect to

electric shock, fire and mechanical hazards

only in accordance with UL 60601-1 and

CAN/CSA C22.2 No. 601.1 84DJ

CLEANING

MR850 Heaterbase Using a damp cloth, clean the humidifier with either of the following: Isopropyl Alcohol, normal dishwashing detergent. Temperature

Probe Always clean before use. Probes can be cleaned with one of the following solutions: Sporicidin

TM

, Cidex

TM

OPA, Sporox

TM

OR sterilize the probe using Ethylene Oxide

sterilization at 55 °C (131 °F), 80 kPa, (allow at least 15 hours for residual ETO to disperse before use). Wipe the airway temperature probes clear of any cleaning residues
before use. Store probes in clean conditions. DO NOT autoclave probes.
CAUTION: DO NOT immerse the heaterbase or temperature probe electrical connections in any liquid.

NOTE:Follow the cleaning agent manufacturers’ instructions carefully. It is the user’s responsibility to qualify any deviations from these procedures, both for disinfecting
efficacy and physical effect on the probe and the heaterbase. See the MR850 technical manual for more detailed instructions on cleaning.

WARNING

Ensure that invasive mode is set for patients
that have bypassed airways.
The use of breathing circuits, chambers or
other accessories which are not approved
by Fisher & Paykel Healthcare may impair
performance or compromise safety.
Ensure that both temperature probe sensors
are correctly and securely fitted. Failure to
do so may result in temperatures in excess
of 41 °C being delivered to the patient.
Ensure maintenance of grounding integrity
by connection to a 'hospital grade'
receptacle. Always disconnect supply before
servicing.
When mounting a humidifier adjacent to a
patient ensure that the humidifier is always
positioned lower than the patient.
The operation of high frequency surgical
apparatus, shortwave or microwave
equipment in the vicinity of the humidifier
may adversely affect its function. If this
occurs, the humidifier should be removed
from the vicinity of such devices.
Do not touch the glass tip of the chamber
temperature probe during use. Keep black
connectors dry at all times.
Visually inspect accessories for
damage before use.

Patents Pending in Principal Countries

of the World



INSTRUCTION SHEET

DO NOT DISCARD

REF 185042343

Rev G 2004-02

MR850 RESPIRATORY HUMIDIFIER

The MR850 respiratory
humidifier is used to
warm and humidify
gases delivered to
patients requiring
mechanical ventilation,
positive pressure
breathing assistance,
or other medical gases.

INTERNATIONAL P O Box 14 348, Panmure, Auckland 1134, New Zealand Tel:+64 9 574 0100 Fax:+64 9 574 0158 Email: info@fphcare.com Web Site: www.fphcare.com
USA Tel: 1800 446 3908 or +1 949 470 3900 Fax:+1 949 470 3933

Advertising
Popular Brands

Fisher Paykel Heater Mr850

Popular manuals